AVALIAÇÃO DA MICROESTRUTURA DE CABEÇAS FEMORAIS USADAS EM IMPLANTES ORTOPÉDICOS APÓS ENSAIO DE SIMULAÇÃO DE DESGASTE

EVALUATION OF FEMORAL HEADS MICROSTRUCTURE USED IN ORTHOPAEDIC IMPLANTS AFTER THE WEAR SIMULATION TEST

DOI

10.5151/2594-5327-21603

Downloads

Baixar Artigo 145 Downloads

Resumo

Neste trabalho estudou-se o impacto que o desgaste causado durante o ensaio de simulação de desgaste (norma ISO 14242-1) pode ter sobre a microestrutura de cabeças femorais fabricadas em aço inoxidável articulando contra polietileno de ultra-alto peso molecular (PEUAPM). A amostra na condição de desgastada foi inicialmente cortada (transversal e sagital), lixada (#180, #360, #600, #1.200 e #1.500) e finalmente polida. A cabeça femoral foi qualitativamente caracterizada quanto a sua composição química por fluorescência de raios X (FRX) e por espectroscopia de energia dispersiva (EDS) a fim de verificar se os elementos químicos estavam de acordo com o exigido na norma ISO 5832-1. A observação do aspecto metalúrgico (grãos austeníticos, tamanho de grão e presença de ferrita delta), conforme exigência da norma ISO 5832-1, foi realizada em microscópio eletrônico de varredura (MEV) e microscópio óptico (MO). A difração de elétrons retro espalhados (EBSD) acoplada ao MEV permitiu mapear a orientação cristalográfica dos grãos das cabeças femorais. Utilizando o MO foi possível determinar o tipo de inclusões, de acordo com a norma ISO 4967. A análise por mapeamento em microssonda EDS acoplada ao MEV foi empregada como técnica auxiliar para a identificação das inclusões. Uma vez que a dureza é um parâmetro que afeta o desgaste, as cabeças femorais foram avaliadas quanto a sua dureza de acordo com a norma ISO 6507-1.

 

In this work, the impact of wear caused during the wear simulation test (according to ISO 14242-1 Standard) on the microstructure of stainless steel femoral heads articulating against ultra-high molecular weight polyethylene (UHMWPE) was evaluated. The worn femoral head (WFH) was initially cut (transversal and sagittal), grinded (#180, #360, #600, #1,200 e #1,500) and finally polished. The chemical composition of the femoral head was qualitatively characterized by X-ray fluorescence (XRF) and energy dispersive spectroscopy (EDS), aiming to determine the chemical elements according to the specified in the ISO 5832-1 Standard. The metallographic observation (austenitic grain, grain size and presence of delta ferrite), according to ISO 5832-1 Standard, was performed by the use of a scanning electronic microscopy (SEM) and optical microscopy (OM). The crystallographic information of the femoral head was obtained by Electron Backscatter Diffraction (EBSD), coupled to SEM. The type of inclusions, according to the ISO 4967 Standard, was determined by MO. The mapping analysis by EDS coupled to SEM was used as auxiliary technique to the identification of inclusions. Once the hardness affects the wear of the femoral heads, it was evaluated according to the ISO 6507-1 Standard.

Palavras-chave

Cabeça femoral; Prótese ortopédica; Desgaste; Microestrutura, Tribologia.

Femoral head; Orthopedic prosthesis; Wear; Microstructure.

Como citar

Lidízio, Leandro Reis; Trommer, Rafael Mello; Peripolli, Suzana Bottega; Cravina, Erica Gravina; Filho, Jorge Trota. AVALIAÇÃO DA MICROESTRUTURA DE CABEÇAS FEMORAIS USADAS EM IMPLANTES ORTOPÉDICOS APÓS ENSAIO DE SIMULAÇÃO DE DESGASTE , p. 4034-4043. In: 67º Congresso da ABM - Internacional / 12º ENEMET - Encontro Nacional de Estudantes de Engenharia Metalúrgica, de Materiais e de Minas, Rio de Jabeiro, 2012.